Abstract:
OBJECTIVE To establish high performance liquid chromatography-evaporative light scattering(HPLC-ELSD) and ion chromatography-inhibition conductivity detection(IC-ICD) analytical method for sodium chloride, 4-hydroxy-1-butanesulfonate and bis-(4-sulfbutyl) ether disodium of sulfonbutyl-β-cyclodextrin.
METHODS HPLC-ELSD method was used for assay and related subsatances test:CD-Screen-IEC column(4 mm×150 mm, 3 μm); gradient elution; flow rate:1.2 mL·min
-1. IC-ICD method was used for assay and related subsatances test:Dionex Ionpac AS19 column(4 mm×250 mm); mobile phase:30 mmol·L
-1 sodium hydroxide solution for elution; flow rate:1.0 mL·min
-1.
RESULTS Two methods had good separation of components. The
S/
N of sodium chloride, sodium 4-hydroxy-1-butane sulfonate and bis-(4-sulfobutyl) ether disodium in HPLC-ELSD method is 0.230, 0.270 and 0.340 ng, respectively. The limit of detection of IC-ICD is 0.041, 0.027 and 0.014 ng.
CONCLUSION There is no obvious difference between HPLC-ELSD and IC-ICD in three impurities content determination of sodium chloride, 4-hydroxy-1-butanesulfonate and bis-(4-sulfobutyl) ether disodium of sulfonbutyl-β-cyclodextrin. It can be selected according to the actual situation. Only according to the detection limit and quantitative limit of the two methods, IC-ICD has higher sensitivity.