DSC法鉴别高风险塑料输液产品

    Identification of Plastic Infusion Products of High Risk by DSC

    • 摘要: 目的 建立高风险塑料输液产品的差示扫描量热分析鉴别方法。方法 初始温度40 ℃,以20 ℃·min-1加热到200 ℃,再以20 ℃·min-1降温到40 ℃,然后以10 ℃·min-1加热到200 ℃,记录DSC图谱。结果 聚丙烯输液瓶、塑料输液容器用聚丙烯组合盖(拉环式,内盖)、三层共挤输液用膜(袋)和五层共挤输液用膜(袋)均显示其特征DSC图谱。结论 该方法准确、灵敏、简便,适用于对高风险塑料输液产品的鉴别。

       

      Abstract: OBJECTIVE To develop a method for identification of plastic infusion products of high risk by differential scanning calorimetry (DSC). METHODS The initial temperature was 40℃, then heated to 200℃ by 20℃·min-1, cooled to 40℃ by 20℃·min-1, then heated to 200℃ by 10℃·min-1, recorded the DSC spectrum.RESULTS PP infusion bottles, combinational closures of PP for plastic infusion containers (with ring-pull) (inner closure), 3-layer co-extrusion film (bag) for infusion, 5-layer co-extrusion film (bag) for infusion showed specific DSC spectrum, respectively. CONCLUSION This method is accurate, sensitive, simple, it can be used for identification of plastic infusion products of high risk.

       

    /