Abstract:
OBJECTIVE To evaluate and verification the dissolution apparatus auto sampling system.
METHODS Employed mechanical qualified dissolution apparatus, evaluate the auto sampling system D, K based on the accuracy(S
S) and precision(S
R) of sampling volume and time of sampling to filter. Verificate the overall performance of dissolution apparatus based on the dissolution of salicylic acid tablets. The ratio of auto to manual in three sampling time points was calculated.
RESULTS D & K auto sampling system sampling accuracy(S
S) were within ±2% and precision(S
R) were ≤ 3%. The sampling time <30 s. It was suggested the D & K auto sampling system were fast, accurate and precisely. The average ratio of auto D to manual were <100%, sometimes were <95%, which the ratio of auto K was in the range of 98%-102%. The results showed that there was a certain deviation in D system, but there was no obvious deviation in K system.
CONCLUSION The auto sampling system should be verificated. It is suggested the acceptal criteria is:sampling time ≤ 30 s, S
S ≤ ±5%, S
R ≤ 5%; Average of R is in the range of 95%-105% and RSD ≤ 5% at time points with ≤ 50% dissolved, which is 97%-103% and RSD ≤ 5% for time points >50%.