Abstract:
OBJECTIVE To establish the new method of detecting amikacin sulfate injection.
METHODS The optimal chemiluminescence of experiment were determined by chemiluminescence coupled with flow injection.Three effects of luminol,potassium permanganate,sodium hydroxide were inverstigated using single-factor and orthogonal experiments.
RESULTS The optimal conditions for amikacin sulfate determanation were as follows:Luminol 5.0×10
-4 mol·L
-1,potassium permanganate 1.0×10
-5 mol·L
-1,sodium hydroxide 0.1 mol·L
-1.Under the optimum experimental conditions,the relative chemiluminescence intensity was linearly related to the concentration of amikacin sulfate from 1.0×10
-8-1.0×10
-6 g·mL
-1(
r=0.991 7).The detection limit of amikacin sulfate was 7.5×10
-9 g·mL
-1(3σ),with RSD of 2.4% for 11 determinations of 1.0×10
-7 g·mL
-1 amikacin sulfate.The recovery rates were >96%.
CONCLUSION The proposed method is simple and accurate.The result is satisfactory.It can be used to determine the content of amikacin sulfate injection.