HPLC同时测定丙戊酸钠口服溶液中丙戊酸钠及防腐剂的含量

    Simultaneous Determination of Valproate Sodium and Preservative in Sodium Valproate Oral Solution by HPLC

    • 摘要: 目的 建立HPLC同时测定丙戊酸钠口服溶液中丙戊酸钠及防腐剂含量的方法。方法 采用Thermo Hypersil BDS C18色谱柱(250 mm×4.6 mm,5 μm),以0.01 mol·L-1磷酸二氢钠(磷酸调节pH为2.3)-乙腈(63∶37)为流动相,流速1.0 mL·min-1,检测波长210,254 nm。结果 丙戊酸钠线性范围为0.99~2.92 mg·mL-1r=0.999 9),平均回收率为98.8%(RSD=0.5%);羟苯甲酯钠线性范围为14.74~58.96 μg·mL-1r=0.999 9),平均回收率为100.8%(RSD=0.4%);羟苯丙酯钠线性范围为10.21~20.41 μg·mL-1r=0.999 9),平均回收率为101.1%(RSD=0.8%)。结论 该方法简单、迅速、可靠,可用于丙戊酸钠口服溶液的检测和质量控制。

       

      Abstract: OBJECTIVE To establish an HPLC method for simultaneous determination of valproate sodium and antiseptic in sodium valproate oral solution. METHODS Thermo Hypersil BDS C18 chromatographic column(250 mm×4.6 mm, 5 μm) was used. The isocratic elution was adopted, with the mobile phase composed of PBS (pH 2.3)-acetonitrile(63:37). The flow rate was 1.0 mL·min-1 and the detective wavelength was 210 and 254 nm. RESULTS The linear range of valproate sodium was 0.99-2.92 mg·mL-1 and the regression coefficient was 0.999 9. The recovery was 98.8%(RSD=0.5%). The linear range of sodium methylparaben was 14.74-58.96 μg·mL-1 and the regression coefficient was 0.999 9. The recovery was 100.8% (RSD=0.4%). The linear range of sodium propylparaben was 10.21-20.41 μg·mL-1 and the regression coefficient was 0.999 9. The recovery was 101.1%(RSD=0.8%). CONCLUSION The method is simple, quick and reliable. It can be used for actual test and quality control of sodium valproate oral solution.

       

    /

    返回文章
    返回