HPLC检测磷酸可待因中的有关物质
胡菲菲1, 胡延雷2, 苗庆柱2, 杨哲洲1, 张福利1
1.中国医药工业研究总院上海医药工业研究院, 上海 201203;2.国药集团工业有限公司, 北京 101301
摘要:
目的 建立HPLC检测磷酸可待因中的有关物质。方法 采用以十八烷基硅烷键合硅胶为填充剂的色谱柱,以缓冲溶液(0.2%磷酸二氢钾1%庚烷磺酸钠的水溶液,磷酸调pH至3.2±0.2)-甲醇(7:3)为流动相A;以流动相A-乙腈(6:4)作为流动相B,采用梯度洗脱;流速1.0 ml·min-1,检测波长245 nm,柱温40℃。结果 该方法主成分与各杂质(A~K)间分离度>1.5。主成分及其各杂质在各自浓度范围内呈现良好的线性关系(r>0.999);平均回收率为96.7%~103.8%,RSD为1.4%~6.0%。结论 该方法操作简单、专属性好、重复性好、准确度高,适用于磷酸可待因的有关物质的测定。
关键词:  磷酸可待因  高效液相色谱法  有关物质
DOI:10.13748/j.cnki.issn1007-7693.2019.17.011
分类号:R917.101
基金项目:
Determination of the Related Substance of Codeine Phosphate by HPLC
Hu Feifei1, Hu Yanlei2, Miao Qingzhu2, Yang Zhezhou1, Zhang Fuli1
1.Shanghai Institute of Pharmaceutical Industry, China State Institute of Pharmaceutical Industry, Shanghai 201203, China;2.China National Pharmaceutical Industry Corporation Ltd., Beijing 101301, China
Abstract:
OBJECTIVE To establish an HPLC method for determination of the related substances in codeine phosphate. METHODS The test was based on an octadecylsilyl silica gel column; the mobile phase A consisted of 0.2% potassium dihydrogen phosphate and 1% sodium heptanesulfonate (pH was adjusted to 3.2±0.2 with phosphoric acid)-methanol(7:3), the mobile phase B consisted of mobile phase A and acetonitrile(6:4), with gradient elution at the flow rate of 1.0 mL·min-1; the detection wavelength was 245 nm; the column temperature was 40℃. RESULTS The separation between the main component of the method and each impurity(A-K) were >1.5. The calibration curves of codeine and known impurities were linear in the self-concentration range; and the average recovery rates(n=9) were in the range of 96.7%-103.8% while RSDs were in th range of 1.4%-6.0%. CONCLUSION The method is improved with good specificity and linearity. The self-control method for proposed main component with correction factors is suitable for the quality control of this drug.
Key words:  codeine phosphate  HPLC  related substance