Determination of Acrylic Acid in Carbomer Copolymer by HPLC
-
Graphical Abstract
-
Abstract
OBJECTIVE To establish an HPLC method for the determination of acrylic acid in carbomer copolymer. METHODS The samples were separated on Sepax GP-C18 column (4.6 mm×250 mm, 5 μm) eluted with potassium dihydrogen phosphate-methanol(80∶20) as mobile phase. The isocratic elution was used at a flow rate of 1.0 mL·min-1, the detection wavelength was 200 nm, the column temperature was 30 ℃. RESULTS The calibration curve was linear in the range of 0.327 5-32.75 μg·mL-1 for acrylic acid with r=0.999 6, the average recovery was 93.02%. CONCLUSION The method is accurate and reliable, and can be applied for the quality control of acrylic acid in carbomer copolymer.
-
-