LI Yan, ZOU Jian, LIU Feng, LI Rui. Determination of Triethylamine in Pramipexole Hydrochloride by Headspace GC[J]. Chinese Journal of Modern Applied Pharmacy, 2014, 31(7): 863-865.
    Citation: LI Yan, ZOU Jian, LIU Feng, LI Rui. Determination of Triethylamine in Pramipexole Hydrochloride by Headspace GC[J]. Chinese Journal of Modern Applied Pharmacy, 2014, 31(7): 863-865.

    Determination of Triethylamine in Pramipexole Hydrochloride by Headspace GC

    • OBJECTIVE To determine the triethylamine in pramipexole hydrochloride by GC. METHODS The residual triethylamine was determined by HS-GC with Agilent-INNOWAX capillary column(30 m×0.32 mm, 0.5 μm) and FID detector. The carrier gas was nitrogen and the flow rate was 4.0 mL·min-1. The temperature of the injection port was maintaining at 250 ℃ and the same of detector. The oven introduced sequential increasing of temperature programing. The initial column temperature was 50 ℃, then raised the temperature to 150℃ at a rate of 10 ℃·min-1, and maintained it at 150 ℃ for 5 min, then raised the temperature to 220 ℃ at a rate of 40 ℃·min-1, and maintained it for 5 min. The headspace oven was set at a temperature of 80 ℃ for 30 min. The solvent was 20% sodium hydroxide solution. The triethylamine was quantified external standard. RESULTS The calibration shows a good linearity with the range of 0.317?12.68 μg·mL-1 for triethylamine. The average recovery was 97.9%(n=9). The precision was 4.18%(n=9). CONCLUSION This method is accurate, reliable and sensitive for the determination of the residual quantity of triethylamine in pramipexole hydrochloride.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return