WANG Wenqiang, CHEN Binbin, WU Hailong, ZHANG Jiaxin, XIE Jiajiang. LC-MS/MS Determination of Methylphenidate in Human Plasma[J]. Chinese Journal of Modern Applied Pharmacy, 2010, 27(10): 942-944.
    Citation: WANG Wenqiang, CHEN Binbin, WU Hailong, ZHANG Jiaxin, XIE Jiajiang. LC-MS/MS Determination of Methylphenidate in Human Plasma[J]. Chinese Journal of Modern Applied Pharmacy, 2010, 27(10): 942-944.

    LC-MS/MS Determination of Methylphenidate in Human Plasma

    • OBJECTIVE To establish an LC-MS/MS method for determination of methylphenidate in human plasma. METHODS 1.0 mL plasma was basified with 40 μL 2 mol·L-1 NaOH and extracted with 4 mL ether. The separation was performed on an Eclipse XDB-C18 column (150 mm×4.6 mm, 5 μm). The mobile phase consisted of methanol and water for gradient elution. Methylphenidate was determined by LC-MS/MS operated in the positive ion mode. RESULTS The limit of quantitation was 0.1 ng·mL-1, and a good linearity was obtained in the range of 1–100 ng·mL-1(r=0.999 1). The RSD of inter-day and intra-day precision were less than 8%. Each analysis was completed within 10 min. CONCLUSION The method is accuracy, sensitive and simple, proved to be suitable for the study of therapeutic drug monitoring of methylphenidate.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return