Determination of Linezolid Content and Related Substances in Linezolid Injection by HPLC
-
Graphical Abstract
-
Abstract
OBJECTIVE To establish an HPLC method for the determination of linezolid and related substances in linezolid injection. METHODS The analytical column was Eclipse XDB-C18, the mobile phase was methanol-water(45∶55), the flow rate was 1.0 mL·min-1. The detection wavelength was 254 nm, and the column temperature was 25 ℃. RESULTS The linear range of calibration curve was 0.5-250 mg·L-1(r=0.999 8), and the limit of detection was 1 ng. The average recovery of three concentrations were 99.3%, 100.9%, 100.5%, respectively. CONCLUSION The method is simple, sensitive and accurate, and can control the quality of the preparation effectively.
-
-