• 首页期刊简介编委会刊物订阅专栏专刊电子刊广告合作联系我们English
引用本文:魏永鸽,徐凯,范明松.1H和19F核磁共振定量法测定塞来昔布含量[J].中国现代应用药学,2020,37(16):1982-1985.
WEI Yongge,XU Kai,FAN Mingsong.Content Determination of Celecoxib by 1H and 19F Quantitative Nuclear Magnetic Resonance[J].Chin J Mod Appl Pharm(中国现代应用药学),2020,37(16):1982-1985.
【打印本页】   【HTML】   【下载PDF全文】   查看/发表评论  【EndNote】   【RefMan】   【BibTex】
←前一篇|后一篇→ 过刊浏览    高级检索
本文已被:浏览 9815次   下载 4481 本文二维码信息
码上扫一扫!
分享到: 微信 更多
1H和19F核磁共振定量法测定塞来昔布含量
魏永鸽1, 徐凯2, 范明松3
1.郑州铁路职业技术学院, 郑州 450052;2.黄河科技学院, 郑州 450005;3.上海雷允上药业有限公司, 上海 201401
摘要:
目的 分别采用1H核磁共振定量法和19F核磁共振定量法测定塞来昔布绝对含量,并将2种方法的测定结果与质量平衡法测定结果进行比较。方法 1H核磁共振定量法以CD3OD为溶剂,塞来昔布δ 6.92处质子峰作为定量峰,马来酸δ 6.33处为内标峰。在弛豫延迟时间10 s,采样时间4.01 s,扫描次数为32条件下采集混合物的氢谱。19F核磁共振定量法以CD3OD为溶剂,4-溴-2-氟-乙酰苯胺为内标,在谱宽240 ppm,中心频率-94.26 ppm,样品扫描次数为16次条件下采集氟谱,计算1H和19F核磁共振定量法的测试结果。结果 1H核磁共振定量法和19F核磁共振定量法测定结果相近,且与质量平衡法测定结果基本一致。结论 核磁共振定量法可以用来测定塞来昔布含量,具有准确、高效的优势。
关键词:  塞来昔布  含量测定  氢核磁共振定量法  氟核磁共振定量法
DOI:10.13748/j.cnki.issn1007-7693.2020.16.011
分类号:R917.101
基金项目:
Content Determination of Celecoxib by 1H and 19F Quantitative Nuclear Magnetic Resonance
WEI Yongge1, XU Kai2, FAN Mingsong3
1.Zhengzhou Railway Vocational & Technical College, Zhengzhou 450052, China;2.Huanghe Science and Technology College, Zhengzhou 450005, China;3.Shanghai LeiYunShang Pharmaceutical Co,. Ltd., Shanghai 201401, China
Abstract:
OBJECTIVE To determine the content of celecoxib by 1H quantitative nuclear magnetic resonance and 19F quantitative nuclear magnetic resonance, and compare the results of the two methods of quantitative nuclear magnetic resonance with mass balance method. METHODS CD3OD was employed as solvent, proton signal peaks of celecoxib at δ 6.92 and δ 6.33 of maleic acid were served as quantitative peaks and inner peak, respectively. The spectra of the mixture were collected at 10 s relaxation delay, 4.01 s acquisition time and 32 scanning times. In the 19F quantitative fluorine nuclear magnetic resonance, CD3OD was used as solvent, 4-bromo-2-fluoroacetanilide was employed as internal standard, the spectra were obtained with spectral width of 240 ppm, transmitter frequency offset was -94.26 ppm and the scanning times was 16. The results of 1H quantitative nuclear magnetic resonance and 19F quantitative nuclear magnetic resonance were calculated. RESULTS The results of 1H quantitative nuclear magnetic resonance and 19F quantitative nuclear magnetic resonance were very close, which were generally consistent with that of the mass balance method. CONCLUSION Quantitative nuclear magnetic resonance can be used for the quantitative determination of celecoxib with the superiority of accurate and effective.
Key words:  celecoxib  content determination  1H quantitative nuclear magnetic resonance  19F quantitative nuclear magnetic resonance
扫一扫关注本刊微信