摘要:
目的 研究环氧乙烷灭菌指示剂D值测定方法。方法 采用部分阴性法测定不同厂家产品的D值,用存活灭杀时间进行验证,两者一致说明部分阴性法测定D值结果有效。若无法验证,则采用残存曲线法重新测定D值,再进行验证。结果 5个样本中,3个用部分阴性法测定的D值可以被存活灭杀时间验证。另2个样本用残存曲线法测定的D值与验证结果一致。结论 D值是评价环氧乙烷灭菌生物指示剂产品质量的关键性技术参数,部分阴性法和残存曲线法的D值测定结果存在差异,为确证D值测定结果的准确性,应采用存活灭杀时间对测定结果进行验证。
关键词:  环氧乙烷  生物指示剂  D值
DOI:10.13748/j.cnki.issn1007-7693.2017.12.006
分类号:R945
基金项目:2014年国家药典委员会标准提高研究项目;浙江省药品接触材料质量控制研究重点实验室(2014E10006)
Study on the Test Method of D Value of Ethylene Oxide Detergent Biological Indicator
CAO Wei, SHEN Hong, LI Jue, WU Xin, LIANG Fayong, LI Chao, SUN Han
Zhejiang Institute for Food and Drug Control, Hangzhou 310052, China
Abstract:
OBJECTIVE To research the test method for D value determination of ethylene oxide sterilization indicator. METHODS The D value of products from different companies were determined by limited Sperman-Karber (LSKP) method and proved by survival and killing time. The consistency of these two results indicated the validity of D value. If D value couldn't be proved by LSKP method, survival-curve method should be selected to examine D value. RESULTS Among the D value of the 5 samples, three of them using LSKP method could be validated by survival and killing time. The results of other two samples examined by survival-curve method were validated by survival and killing time successfully. CONCLUSION D value is the key technical parameter to evaluate the product quality of ethylene oxide sterilization biological indicator. The results of D value examined by LSKP method and survival-curve method show the difference. Survival and killing time is recommended to be applied in validation to assure the accuracy of D value.
Key words:  ethylene oxide  biological indicator  D value